Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 38060 |
| Brand Name: | Hamilton/47244/Ventilators |
| Model: | HAMILTON-C6 Part number: 160021 |
| Affected: | Serial numbers: 17196, 17357, 17358, 17361, 17362, 17363 |
| Software version: | |
| Recalling Organisation: | InterMed Medical Ltd 71 Apollo Drive Albany AUCKLAND 1311 |
| Contact Information: | 0800 333 444 |
| Manufacturer: | Hamilton Medical AG |
| Issue: | Certain ventilators with serial numbers are equipped with a blower module that is prone to reach the end of its service life earlier than expected. In some cases, the ventilator was alarming with technical events or technical faults (e.g. TE 231009 blower speed low or TF 431001 blower fault), indicating a technical fault of the blower. The issue might occur even if the pre-operative test of the ventilator passed. If the ventilator alarms with a technical fault during active ventilation with a patient, the device might transition to ambient state, resulting in loss of mechanical ventilation and requiring immediate provision of alternative ventilation. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 25/11/2026 |