Database updated: 27 November 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 38039 |
| Brand Name: | Corpuls 3 Defib including Accessories and Upgrades |
| Model: | Corpuls 3 and Corpuls 3T |
| Affected: | Product Code: CP-04100, CP-04101 - Multiple serial numbers |
| Software version: | |
| Recalling Organisation: | Universal Specialities Ltd trading as USL Medical 494 Rosebank Road Avondale AUCKLAND 1026 |
| Contact Information: | 09 829 0595 |
| Manufacturer: | GS Elektromedizinische GmbH |
| Issue: | The manufactuere has identified a potential correlation between printing a resting ECG and a temporary interruption of the monitor's Bluetooth connection, which may result in an unexpected, brief interruption of the display. A temporary interruption of the display may result in users temporarily losing access to displayed patient information and being unable to operate the system for 5–25 seconds. Particularly in emergency situations, this may lead to an interruption or delay in life-saving treatment and, in the worst case, may result in serious adverse health consequences for patients. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 23/11/2026 |