Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 38031 |
| Brand Name: | Stryker Triathlon Hinge Trays |
| Model: | Catalogue number: 6543-6-506, 6543-6-507, 6543-6-519, 6543-6-514 |
| Affected: | Lot number: 19114556, 200009859, 200009873, 20447278, 19114564, 200009912, 19114388, 70945430, 70945448 |
| Software version: | |
| Recalling Organisation: | Stryker New Zealand Ltd 515 Mount Wellington Highway Mount Wellington AUCKLAND 1060 |
| Contact Information: | 0800 787 9287 |
| Manufacturer: | Stryker Howmedica Osteonics |
| Issue: | The manufacturer has identified that the reprocessing instructions provided in LSTPI-B specify a sterilisation dry time of 30 minutes for the trays. However, these trays require a sterilisation dry time of 40 minutes when sterilised in blue wrap. As a result, when blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilisation. This action is being conducted to notify customers of the hazards and harms associated with the reduced dry time. Products continue to be safe to use when reprocessed in accordance with the updated instruction of 40-minute dry time when sterilised in blue wrap. No product return is necessary. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 23/11/2026 |