Detail of Recall

Type of Product:Device
Medsafe Reference:38031
Brand Name:Stryker Triathlon Hinge Trays
Model:Catalogue number: 6543-6-506, 6543-6-507, 6543-6-519, 6543-6-514
Affected:Lot number: 19114556, 200009859, 200009873, 20447278, 19114564, 200009912, 19114388, 70945430, 70945448
Software version:
Recalling Organisation:Stryker New Zealand Ltd
515 Mount Wellington Highway
Mount Wellington
AUCKLAND 1060
Contact Information:0800 787 9287
Manufacturer:Stryker Howmedica Osteonics
Issue:The manufacturer has identified that the reprocessing instructions provided in LSTPI-B specify a sterilisation dry time of 30 minutes for the trays. However, these trays require a sterilisation dry time of 40 minutes when sterilised in blue wrap. As a result, when blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilisation.
This action is being conducted to notify customers of the hazards and harms associated with the reduced dry time. Products continue to be safe to use when reprocessed in accordance with the updated instruction of 40-minute dry time when sterilised in blue wrap. No product return is necessary.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/11/2026