Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 38024 |
| Brand Name: | Infinity Ammonia Reagent for Beckman Coulter AU Chemistry Analyzers |
| Model: | Infinity Ammonia Reagent |
| Affected: | Multiple Lots |
| Software version: | |
| Recalling Organisation: | Beckman Coulter NZ Ltd Suite 2, Level 4 30 St Benedicts Street Newton AUCKLAND |
| Contact Information: | 0800 446 109 |
| Manufacturer: | Fisher Diagnostics |
| Issue: | The manufacturer is initiating an urgent Recall notice for the product Infinity Ammonia Reagent for Beckman Coulter AU Chemistry Analyzers because the manufacturer have received complaints from customers reporting approximately 6% to 15% low calibration bias after switching ammonia standard lots and attributed the bias to the calibrators. This observed issue applies to customers who purchased the product. |
| Recall action type: | Recall |
| Recall action: | Product to be destroyed |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 19/11/2026 |