Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 38002 |
| Brand Name: | ROTAFLOW System (Bypass system) |
| Model: | Item: 701051697 RotaFlow English/UK UK-Plug ICU |
| Affected: | All products currently on the market are affected. 2005 through 2020-12-31 |
| Software version: | |
| Recalling Organisation: | Getinge Australia Pty Ltd (NZ) Unit D, Ground Floor, Building A, Millenium Centre 600 Great South Rd Ellerslie Auckland 1051 |
| Contact Information: | 09 272 9039 |
| Manufacturer: | Maquet Cardio Pulmonary GmbH |
| Issue: | During internal validation of new replacement batteries, the manufacturer identified a problem on the ROTAFLOW Console. This problem can cause the device to stop battery charging after the battery operation mode has been utilised. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 18/11/2026 |