Database updated: 20 November 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:38002
Brand Name:ROTAFLOW System (Bypass system)
Model:Item: 701051697 RotaFlow English/UK UK-Plug ICU
Affected:All products currently on the market are affected. 2005 through 2020-12-31
Software version:
Recalling Organisation:Getinge Australia Pty Ltd (NZ)
Unit D, Ground Floor, Building A, Millenium Centre
600 Great South Rd
Ellerslie
Auckland 1051
Contact Information:09 272 9039
Manufacturer:Maquet Cardio Pulmonary GmbH
Issue:During internal validation of new replacement batteries, the manufacturer identified a problem on the ROTAFLOW Console. This problem can cause the device to stop battery charging after the battery operation mode has been utilised.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:18/11/2026
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /