Detail of Recall

Type of Product:Device
Medsafe Reference:37990
Brand Name:BD Connecta Stopcock with Extension Tube
Model:Product code: 394995 BD Connecta™ Stopcock with Extension Tube - white 16 cm
Affected:Lot number: 5042456, 5062802, 5073966, 5073969, 5098804, 5120823
Software version:
Recalling Organisation:Becton Dickinson Ltd
Unit 4 / 110 Mays Road
Onehunga
Auckland 1061
Contact Information:0800 572 468
Manufacturer:BD Switzerland Sarl
Issue:The manufacturer has identified that specific lot numbers 394995 of the BD Connecta Stopcock with Extension Tube product may have been manufactured with an inadequate amount of solvent which could result in separation of the tubing from the male luer component.
Separation of the Connecta™ Stopcock with Extension Tube may occur during use, resulting in a compromised fluid pathway. This may result in leakage with the potential consequences including under-infusion or disruption of prescribed therapy, exposure to potentially hazardous fluids, infection or air embolism.
The separation may be apparent upon initiation of infusion, or during standard patient care at which point the healthcare professional should discontinue use and replace the affected stopcock extension set.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:18/11/2026