Detail of Recall

Type of Product:Device
Medsafe Reference:37983
Brand Name:Brainlab Cranial EM
Model:Navigation Software Cranial EM utilising Universal Atlas 6.0
Affected:software versions: 1.0, 1.1.1, 1.1.2, 1.1.3, 1.1.4, 1.1.5.
Software version:
Recalling Organisation:Avania Pty Ltd
Level 2
142 Broadway
Newmarket
Auckland 1023
Contact Information:
Manufacturer:Brainlab SE
Issue:There is a potential safety problem when using Navigation Software Cranial EM in combination with the software submodule Universal Atlas 6.0 on neonates (under 29 days old) and infants (29 days to under 2 years old): The Surface Matching registration to navigation might not be possible or
lead to unusable results for these patients.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/11/2026