Database updated: 20 November 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 37983 |
| Brand Name: | Brainlab Cranial EM |
| Model: | Navigation Software Cranial EM utilising Universal Atlas 6.0 |
| Affected: | software versions: 1.0, 1.1.1, 1.1.2, 1.1.3, 1.1.4, 1.1.5. |
| Software version: | |
| Recalling Organisation: | Avania Pty Ltd Level 2 142 Broadway Newmarket Auckland 1023 |
| Contact Information: | |
| Manufacturer: | Brainlab SE |
| Issue: | There is a potential safety problem when using Navigation Software Cranial EM in combination with the software submodule Universal Atlas 6.0 on neonates (under 29 days old) and infants (29 days to under 2 years old): The Surface Matching registration to navigation might not be possible or lead to unusable results for these patients. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 17/11/2026 |