Database updated: 27 November 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:37961
Brand Name:EYE-LIGHT (MDR)
Model: EYE-LIGHT
Affected:All serial numbers
Software version:
Recalling Organisation:Kevin Grundy (IBD) Pty Ltd
C/- NsaTax Ltd
The Keystone Building, Level 6
3 City Road
Auckland
Contact Information:
Manufacturer:Espansione Marketing SpA
Brand Name:MY MASK-E
Model:MY MASK-E
Affected:All serial numbers
Software version:
Recalling Organisation:Kevin Grundy (IBD) Pty Ltd
C/- NsaTax Ltd
The Keystone Building, Level 6
3 City Road
Auckland
Contact Information:
Manufacturer:Espansione Marketing SpA
Brand Name:EYE-LIGHT
Model:Eye-Light
Affected:All serial numbers
Software version:
Recalling Organisation:Kevin Grundy (IBD) Pty Ltd
C/- NsaTax Ltd
The Keystone Building, Level 6
3 City Road
Auckland
Contact Information:
Manufacturer:Espansione Marketing SpA
Issue:Investigation found that use of the EYE-MASK LED mask in patients with certain anatomical or ophthalmic conditions may reduce the required 5–10 mm distance between the skin and the LEDs, resulting in excessive localised heating. These conditions include exophthalmos or proptosis, periocular
oedema and dermatochalasis.
The manufacturer is updating the user manuals for EYE-LIGHT, MY MASK-E and EYE-LIGHT to include additional warnings for use in patients with these conditions.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Healthcare Professional
Date Commenced:24/11/2026
Hide menus