| Type of Product: | Device |
| Medsafe Reference: | 37959 |
| Brand Name: | Transpac IV Dual Cable, Dual IBP Adapter |
| Model: | Dual IBP Adapter |
| Affected: | Product model number: 989803199741 |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medizin Systeme Boblingen GmbH |
| Issue: | The manufacturer has identified that the Dual IBP Adapter may not provide adequate defibrillation protection when used with non-defibrillation-proof transducers that are not listed as compatible in the IFU. In the event of defibrillation, energy may be partially diverted through the IBP system rather than the patient, potentially creating a hazardous situation. The risk is particularly relevant when catheter-tip transducers are used, as these are typically not defibrillation-proof. The adapter design does not prevent connection to incompatible transducers. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 12/11/2026 |