Detail of Recall

Type of Product:Device
Medsafe Reference:37959
Brand Name:Transpac IV Dual Cable, Dual IBP Adapter
Model:Dual IBP Adapter
Affected:Product model number: 989803199741
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medizin Systeme Boblingen GmbH
Issue:The manufacturer has identified that the Dual IBP Adapter may not provide adequate defibrillation protection when used with non-defibrillation-proof transducers that are not listed as compatible in the IFU. In the event of defibrillation, energy may be partially diverted through the IBP system rather than the patient, potentially creating a hazardous situation. The risk is particularly relevant when catheter-tip transducers are used, as these are typically not defibrillation-proof. The adapter design does not prevent connection to incompatible transducers.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/11/2026