Type of Product:Device
Medsafe Reference:37933
Brand Name:DxC 500i AU Clinical Analyzer
Model:With ISE - AU w/ISE and Without ISE -AU w/out ISE - software 1.5.0 and below
Affected:
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Inc
Issue:The manufacturer has identified an issue that may affect the reporting of Clinical Chemistry test results. When a laboratory selects a reporting unit other than the system regional default unit for a Clinical Chemistry test and multiple tests are ordered under the same Sample ID, or multiple replicates of the same test under the same Sample ID, the software may incorrectly apply the unit conversion factor more than once. This may result in an erroneous reported value.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:10/11/2026