Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 37933 |
| Brand Name: | DxC 500i AU Clinical Analyzer |
| Model: | With ISE - AU w/ISE and Without ISE -AU w/out ISE - software 1.5.0 and below |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Beckman Coulter NZ Ltd Suite 2, Level 4 30 St Benedicts Street Newton AUCKLAND |
| Contact Information: | 0800 446 109 |
| Manufacturer: | Beckman Coulter Inc |
| Issue: | The manufacturer has identified an issue that may affect the reporting of Clinical Chemistry test results. When a laboratory selects a reporting unit other than the system regional default unit for a Clinical Chemistry test and multiple tests are ordered under the same Sample ID, or multiple replicates of the same test under the same Sample ID, the software may incorrectly apply the unit conversion factor more than once. This may result in an erroneous reported value. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 10/11/2026 |