Detail of Recall

Type of Product:Device
Medsafe Reference:37865
Brand Name:RIDA Setup 3.01.05
Model:RIDA Xplore System (96521) and HighPlex System (91501)
Affected:
Software version:
Recalling Organisation:AusDiagnostics NZ Ltd
156B Bush Road
Albany
Auckland 0632
Contact Information:
Manufacturer:AusDiagnostics Pty Ltd
Issue:A software defect was identified in RIDA® Setup v3.01.05, affecting the RIDA®Xplore System (96521) and HighPlex System (91501) The defect occurs only when all the following conditions are met:
· Samples are entered manually;
· The POSCON option is selected; and
· The positive control is placed in any position other than the last position in the run.
Under these conditions, the software repositions a sample without updating the corresponding sample identifier, resulting in a mismatch between the physical sample position and the assigned sample label. This creates the potential for incorrect sample identification and assignment of results to the wrong specimen.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:4/11/2026