Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 37844 |
| Brand Name: | Anaesthesia Work Stations |
| Model: | Carestation 850 and 850c Anesthesia Systems |
| Affected: | Affected Replaceable Unit part: 5830520-S - SELECTOR VALVE ASSEMBLY |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | Datex-Ohmeda Inc |
| Issue: | The manufacturer has become aware that certain Carestation 850 and 850c Anesthesia Systems, can deliver less anesthetic agent than the user selected concentration setting. Reduced delivery of anesthetic agent can only occur when two vaporizers (either Sevoflurane or Isoflurane) are installed on the system, and the left mounted vaporizer is selected for anesthetic agent administration. This issue can be detected during preoperative checkout or while in clinical use. In the unlikely case this issue occurs during clinical use and the audible alarm and visual alarm messages are not noticed, it could result in patient movement during surgery and/or intraoperative awareness. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 2/11/2026 |