Type of Product:Device
Medsafe Reference:37844
Brand Name:Anaesthesia Work Stations
Model:Carestation 850 and 850c Anesthesia Systems
Affected:Affected Replaceable Unit part: 5830520-S - SELECTOR VALVE ASSEMBLY
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:Datex-Ohmeda Inc
Issue:The manufacturer has become aware that certain Carestation 850 and 850c Anesthesia Systems, can deliver less anesthetic agent than the user selected concentration setting. Reduced delivery of anesthetic agent can only occur when two vaporizers (either Sevoflurane or Isoflurane) are installed on the system, and the left mounted vaporizer is selected for anesthetic agent administration.
This issue can be detected during preoperative checkout or while in clinical use.
In the unlikely case this issue occurs during clinical use and the audible alarm and visual alarm messages are not noticed, it could result in patient movement during surgery and/or intraoperative awareness.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:2/11/2026