Detail of Recall

Type of Product:Device
Medsafe Reference:37775
Brand Name:Hologic - Horizon W X-Ray Bone DXA Systems WI
Model:Part number: Horizon-W
Affected:Lot number: 308911M
Software version:
Recalling Organisation:Hologic (Australia and New Zealand) Pty Ltd
c/o Logical Freight Solutions NZ Ltd
Unit 8/113 Pavilion Drive
Airport Oaks
Auckland 2022
Contact Information:
Manufacturer:Hologic Inc
Issue:The Horizon WI systems manufactured from 11 December 2024 to 16 January 2025 were shipped without the validated VGA cables. Some installations were completed using VGA cables that may not meet Hologic's applicable performance and safety standards for this system. The impacted units may present a potential risk of elevated Electromagnetic Interference (EMI) to users and/or patients near the affected Horizon DXA systems. The impact can range from user dissatisfaction to delay in treatment and presents an elevated risk of interference to the population of individuals with active, implantable medical devices.
Recall action type:Correction
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:27/10/2026