Detail of Recall

Type of Product:Device
Medsafe Reference:37767
Brand Name:NOXBOXi Nitric Oxide Monitor and Delivery System
Model:Product Code - NOXBOX-I
Affected:Serial number: NI103206, NI100931, NI101157, NI101195, NI103009, NI100792
Software version:
Recalling Organisation:Keyport Ltd
431 Te Rapa Road
HAMILTON
Contact Information:
Manufacturer:NOxBOX Ltd
Issue:The manufacturer has identified a manufacturing issue affecting certain oxygen sensors. The issue may cause oxygen sensors to intermittently display irregular oxygen concentrations or cause the NOxBOXi device to fail the required oxygen sensor zero calibration. This sensor failure may be
intermittent and may not be detected immediately following manufacture or during pre-use system and setup checks. Based on the design of the NOxBOXi device, the oxygen sensor is used solely to monitor the oxygen level in the sampled gas and is complimentary to other oxygen monitoring that might be used on a patient. It is not used by the device to automatically adjust or control oxygen delivery, nor is the
monitored value used to control iNO delivery during Intelligent Delivery mode.
Recall action type:Correction
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/10/2026