Detail of Recall

Type of Product:Device
Medsafe Reference:37433
Brand Name:NAV- Passive Spheres
Model:NDI Passive Sphere – Model No. 8801071, 8801074, 8801075 (specific lots)
Affected:
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Northern Digital Inc
Brand Name:CST - NAV - Surgical Procedure Kit, general purpose
Model:Medtronic Biopsy Needle Kit – Model No. 9733068 (specific lots)
Affected:
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Navigation Inc
Issue:The manufacturer c has identified variability in the reflective material applied to a limited number of NDI Sterile Passive Sphere lots. The reflective coating is used to reflect infrared light from optical tracking cameras, allowing optical navigation systems to detect and track instruments during surgical procedures.
The affected spheres may not be detected or tracked as intended by compatible navigation systems. This may result in instrument verification failures, intermittent or complete loss of tracking, tracking status errors, or reduced navigation accuracy. The variation in reflective material is not visually detectable and affected spheres cannot be reliably distinguished from unaffected spheres through inspection.
This variability also affects specific lots of the Medtronic Passive Biopsy Needle as the needle includes two NDI passive spheres built within the instrument.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Healthcare Professional
Date Commenced:1/9/2026