Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 37345 |
| Brand Name: | [MDR]Philips Patient Information Centre iX PIC iX |
| Model: | Philips Patient Information Centre iX (PIC iX) 4.x systems |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medizin Systeme Boblingen GmbH |
| Issue: | The manufacturer received reports that some PIC iX systems were frozen and could not be used. It was determined that failure of the watchdog may cause the system to be in an unknown state and therefore potentially prevent alarms from being generated or logged. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 24/8/2026 |