Detail of Recall

Type of Product:Device
Medsafe Reference:37345
Brand Name:[MDR]Philips Patient Information Centre iX PIC iX
Model:Philips Patient Information Centre iX (PIC iX) 4.x systems
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medizin Systeme Boblingen GmbH
Issue:The manufacturer received reports that some PIC iX systems were frozen and could not be used. It was determined that failure of the watchdog may cause the system to be in an unknown state and therefore potentially prevent alarms from being generated or logged.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:24/8/2026