Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 37312 |
| Brand Name: | Hemospray Endoscopic Hemostat |
| Model: | HEMOSPRAY ENDOSCOPIC HEMOSTAT 7FR and HEMOSPRAY ENDOSCOPIC HEMOSTAT 10FR |
| Affected: | Product code: G56572, G21049 |
| Software version: | |
| Recalling Organisation: | Obex Medical Ltd Level 3, 109 Carlton Gore Road Newmarket Auckland 1023 |
| Contact Information: | 0800 656 539 |
| Manufacturer: | Wilson-Cook Medical Inc |
| Issue: | The updated Hemospray Instructions for Use (IFU) will include new guidance to address endoscope adherence to the distal end of the endoscope and inability to spray powder. Instructions will be added to mitigate the occurrence of powder adhering to the distal end of the endoscope. This condition can cause the endoscope to adhere to tissue, potentially resulting in difficulty maneuvering or removing the endoscope. The Potential Complications section of the Hemospray IFU will be updated regarding powder adherence to the distal end of the endoscope. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 18/8/2026 |