Detail of Recall

Type of Product:Device
Medsafe Reference:37293
Brand Name:STAAR Surgical Company Stella 2.0 Implant Orientation Diagram (IOD)
Model:MYOPIC TORIC EVO VTICMO12.6-10.5
Affected:Serial number - T2717072
Software version:
Recalling Organisation:Device Technologies New Zealand
47 Arrenway Drive
Albany
AUCKLAND 0632
Contact Information:09 913 2000
Manufacturer:Staar Surgical AG
Issue:The manufacturer has become aware that a limited number of IODs generated have the risk of an IOD with an incorrect axis value that could be clinically significant. Specifically, per the review of data in Stella 2.0, the IOD generated for a device you had ordered has a different axis placement than what had been reserved in Stella 2.0.
Investigation points to an issue arising from migration of data from Stella 1.0 to Stella 2.0 associated with Made to Order (MTO) Toric ICLs. This data migration occurred prior to release of Stella 2.0 on April 13, 2026. The following three failure modes could result in this error:
1. Different IODs could be generated for the same serial number and one of the IODs from the record migrated from Stella 1.0 to Stella 2.0 displays an incorrect axis value, resulting in an incorrect rotation diagram.
2. Stella 2.0 could incorrectly generate an IOD diagram with no lens rotation required.
3. A number of migrated records were converted with null axis values.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/8/2026