Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 37289 |
| Brand Name: | Getinge Steam Sterilizer |
| Model: | HS 6610 AR1, HS 6613 ER-1, HS 6613 AR-2, HS 6613 AC-1, HS 6610 AC-1, HS 6610 EC-1, HS 6610 AR-1, HS 6613 AR-1 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Getinge Australia Pty Ltd (NZ) Unit D, Ground Floor, Building A, Millenium Centre 600 Great South Rd Ellerslie Auckland 1051 |
| Contact Information: | 09 272 9039 |
| Manufacturer: | Getinge Sterilization AB |
| Brand Name: | LTSF Sterilizer |
| Model: | Getinge Steam Sterilizer Door |
| Affected: | HS 5510 AC-1, HS 6613 DR-2, HS 5510 EC-1, HS 6610 ER-1, HS 6613T AR-1, HS 6613T AR-2, HS 6613 ER-2, HS 5510 ER-1 |
| Software version: | |
| Recalling Organisation: | Getinge Australia Pty Ltd (NZ) Unit D, Ground Floor, Building A, Millenium Centre 600 Great South Rd Ellerslie Auckland 1051 |
| Contact Information: | 09 272 9039 |
| Manufacturer: | Getinge Sterilization AB |
| Issue: | The manufacturer received a report of a user injury involving the autoclave door. The sterilization technician activated the door lock while a basket was obstructing the door's movement. Rather than reopening the door and repositioning the basket, the technician attempted to manually adjust the basket while the door was closing. When the basket became dislodged, the door closed and resulted in a serious crushing injury to the technician's hand. During further investigation, a second complaint involving a similar device was also identified. |
| Recall action type: | Alert |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 17/8/2026 |