Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:37257
Brand Name:Howmedica Osteonics Corporation - Trident 0 Degree Constrained Acetabular Insert
Model:Multiple affected lots
Affected:
Software version:
Recalling Organisation:Stryker New Zealand Ltd
515 Mount Wellington Highway
Mount Wellington
AUCKLAND 1060
Contact Information:0800 787 9287
Manufacturer:Stryker Howmedica Osteonics
Brand Name:Constrained Liner
Model:Multiple affected lots
Affected:
Software version:
Recalling Organisation:Stryker New Zealand Ltd
515 Mount Wellington Highway
Mount Wellington
AUCKLAND 1060
Contact Information:0800 787 9287
Manufacturer:Stryker Howmedica Osteonics
Brand Name:Howmedica Osteonics Corporation - UHR Universal Head Bipolar Component
Model:Multiple affected lots
Affected:
Software version:
Recalling Organisation:Stryker New Zealand Ltd
515 Mount Wellington Highway
Mount Wellington
AUCKLAND 1060
Contact Information:0800 787 9287
Manufacturer:Stryker Howmedica Osteonics
Issue:The manufacturer has become aware that the devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five (5) years.
Industry guidance for storage duration of UHMWPE raw material used in the manufacture of implant bearing/articulating material of prosthetic joints is less than five (5) years. Product manufactured using UHMWPE raw material beyond the recommended storage duration has the potential for elevated levels of oxidation. Oxidation of UHMWPE material can impact material properties, potentially resulting in increased wear.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/8/2026