Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:37212
Brand Name:Dilator, vascular
Model:PIK dilator set L#
Affected:Part number: 701054427. All serial numbers
Software version:
Recalling Organisation:Getinge Australia Pty Ltd (NZ)
Unit D, Ground Floor, Building A, Millenium Centre
600 Great South Rd
Ellerslie
Auckland 1051
Contact Information:09 272 9039
Manufacturer:Maquet Cardio Pulmonary GmbH
Brand Name:Catheterization kit, intravenous
Model:PIK 100#Insertion Kit, Guidewire 100 cm and PIK 150#Insertion Kit, Guidewire 150 cm
Affected:Part number:701047384 and 701047385. All serial numbers
Software version:
Recalling Organisation:Getinge Australia Pty Ltd (NZ)
Unit D, Ground Floor, Building A, Millenium Centre
600 Great South Rd
Ellerslie
Auckland 1051
Contact Information:09 272 9039
Manufacturer:Maquet Cardio Pulmonary GmbH
Issue:The manufacturer became aware of this problem during retrospective review of historical packaging verification documentation. It was identified that there have been failures in sterile barrier integrity verification testing. As these failures were not previously addressed in accordance with packaging verification requirements, the packaging may not consistently maintain sterile barrier integrity.
Therefore, this Product Recall is limited to all products of the PIK line within shelf-life.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/8/2026