Detail of Recall

Type of Product:Device
Medsafe Reference:37200
Brand Name:[MDR] DigitalDiagnost, Digital Diagnost C90, 7300C
Model:Model number 712034: DigitalDiagnost C90 High Performance and Model number 712035: DigitalDiagnost C90 Flex/Value/Chest/ER
Affected:Software version - 1.1.7V20.4270.008
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems DMC GmbH
Issue:The manufacturer has identified multiple software problems affecting specific DXR systems. The problems have no associated safety impact but may affect the performance of the equipment.
1. The system may not enter a ready state for wall stand exposures when the Ceiling Suspension Motorised (CSM) Arm Swing is not in the locked position, preventing certain examinations from being performed as intended.
2. The Tube Control Handle (TCH) may intermittently restart
multiple times during start-up, resulting in delays before the system is available for clinical use.
3. Systems without a valid 2D camera licence may display the live camera image on the Acquired Work Station (AWS).
4. Following a language change, the TCH display may not immediately synchronise with the selected system language after restart.
5. The CSHeightOffset calculation used by the Smart Collimation Thorax (SCT) feature may be less accurate for source-to-image
distances greater than 150 cm.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:4/8/2026