Detail of Recall

Type of Product:Device
Medsafe Reference:37180
Brand Name:[MDR] Azurion
Model:Philips Azurion R3.1 Systems
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:The manufacturer has become aware of an anomaly with the Philips Azurion R3.1 Systems.
Issue 1 - System remains in continuous restart mode after the start-up Due to a software issue, the License Manager – which verifies installed licenses and enables optional functionality in the Azurion system based on the available licenses – expands the Windows registry database at every system start or restart.
Issue 2 - Longitudinal Position Error – Applicable only to Azurion systems with Poly-G3 frontal stand When requesting a movement of the frontal stand in longitudinal direction, the Azurion system may detect a mismatch between the expected longitudinal (set-) position and the measured (actual) position of the stand. This mismatch is caused by a glitch in the longitudinal position value provided by the position sensing potentiometer.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/8/2026