Detail of Recall

Type of Product:Device
Medsafe Reference:37157
Brand Name:Intra-Aortic Balloon Balloon pump
Model:Product: Intra-Aortic Balloon CS100 IABP, CS300 IABP
Affected:
Software version:
Recalling Organisation:Getinge Australia Pty Ltd (NZ)
Unit D, Ground Floor, Building A, Millenium Centre
600 Great South Rd
Ellerslie
Auckland 1051
Contact Information:09 272 9039
Manufacturer:Datascope Corporation
Issue:The manufacturer has become aware of instances where the device did not provide the expected runtime during some portable IABP uses CS100/CS300 devices. The manufacturer recently completed additional testing on the CS100/CS300 batteries and identified the battery runtime and cycle specifications contained in the devices' Instructions for Use (IFU) were incorrect and need to be updated. No other CS100/CS300 IABP performance deficiencies were identified.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/7/2026