Detail of Recall

Type of Product:Device
Medsafe Reference:37151
Brand Name:Corpuls 3 Defib including Accessories and Upgrades
Model:corpuls3 SLIM/SLIM TOUCH Defibrillator Module
Affected:Product codes: 04301, 04302. All serial numbers
Software version:
Recalling Organisation:Universal Specialities Ltd trading as USL Medical
494 Rosebank Road
Avondale
AUCKLAND 1026
Contact Information:09 829 0595
Manufacturer:GS Elektromedizinische GmbH
Issue:The manufacturer has identified a potential problem associated with the magnetic charging interface (MagCode) of the corpuls3-SLIM/SLIM Touch Modules.
Increased contact resistance may develop within the magnetic charging interface of the corpuls3-SLIM/SLIM Touch. This may occur due to:
· material combinations with differing thermal expansion characteristics;
· electrochemical effects such as oxidation or corrosion arising from different material potentials
· wear, contamination, or aging of the contact surfaces.
As a result, the following effects may occur:
· localised heat generation at the charging interface;
· visible discolouration, deformation, or scorching of the connector or cable;
· elevated surface temperature at the charging contact.
This problem is limited to the charging interface only and does not affect monitoring, therapy, or alarm functions.
Recall action type:Correction
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:27/7/2026