Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 37151 |
| Brand Name: | Corpuls 3 Defib including Accessories and Upgrades |
| Model: | corpuls3 SLIM/SLIM TOUCH Defibrillator Module |
| Affected: | Product codes: 04301, 04302. All serial numbers |
| Software version: | |
| Recalling Organisation: | Universal Specialities Ltd trading as USL Medical 494 Rosebank Road Avondale AUCKLAND 1026 |
| Contact Information: | 09 829 0595 |
| Manufacturer: | GS Elektromedizinische GmbH |
| Issue: | The manufacturer has identified a potential problem associated with the magnetic charging interface (MagCode) of the corpuls3-SLIM/SLIM Touch Modules. Increased contact resistance may develop within the magnetic charging interface of the corpuls3-SLIM/SLIM Touch. This may occur due to: · material combinations with differing thermal expansion characteristics; · electrochemical effects such as oxidation or corrosion arising from different material potentials · wear, contamination, or aging of the contact surfaces. As a result, the following effects may occur: · localised heat generation at the charging interface; · visible discolouration, deformation, or scorching of the connector or cable; · elevated surface temperature at the charging contact. This problem is limited to the charging interface only and does not affect monitoring, therapy, or alarm functions. |
| Recall action type: | Correction |
| Recall action: | Product to be exchanged |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 27/7/2026 |