Detail of Recall

Type of Product:Device
Medsafe Reference:37147
Brand Name:[MDR] PageWriter TC35 Cardiograph; CW 5000; CW7000
Model:Cardiac Workstation 5000
Affected:Product Number - 860439
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medizin Systeme Boblingen GmbH
Issue:The manufacturer received multiple complaints that the Cardiac Workstation may intermittently display an inaccurate, often elevated, native heart rate. While clinicians were generally able to identify the discrepancy because it did not match the patient's clinical presentation, the issue can affect the accuracy of the entire ECG report, including waveform data, measurements, and interpretive statements. The problem is linked to the Wired PIM accessory and may occur, in rare cases, when the Cardiac Workstation exits Standby mode and fails to correctly set the ECG sampling rate. This can result in inaccurate ECG waveform display and interpretation.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:27/7/2026