Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:37127
Brand Name:Surgical Procedure Kit, general purpose
Model:St Lumbar Puncture Kit 2
Affected:Lot number - 8426525
Software version:
Recalling Organisation:Multigate NZ Ltd
621 Rosebank Rd
Avondale
Auckland 1026
Contact Information:
Manufacturer:Multigate Medical Products Pty Ltd
Issue:The manufacturer has identified an observed problem in how Port C of the BD Connecta™ Stopcocks operates during use. Reports received indicate that when using polypropylene plastic luer lock syringes, the connection at Port C may over thread the syringe and continue to spin. This behaviour has led some clinical users to perceive the connection as unstable.
Recall action type:Alert
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/7/2026