Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 37115 |
| Brand Name: | Hoffmann II Micro External Fixation System |
| Model: | Connecting Rod Hoffmann® II Micro 3 x 30mm, 40mm, 50mm, 60mm, 90mm, 120mm, 150mm |
| Affected: | All Lots |
| Software version: | |
| Recalling Organisation: | Stryker New Zealand Ltd 515 Mount Wellington Highway Mount Wellington AUCKLAND 1060 |
| Contact Information: | 0800 787 9287 |
| Manufacturer: | Stryker GmbH |
| Issue: | The manufacturer has confirmed that cracking of the Hoffmann® II Carbon Connecting Rods may occur following cleaning and sterilisation. Contrary to the original product labelling, the part is not suitable for reprocessing, and cleaning and sterilisation should not be performed, as this may result in cracks developing. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 21/7/2026 |