Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:37115
Brand Name:Hoffmann II Micro External Fixation System
Model:Connecting Rod Hoffmann® II Micro 3 x 30mm, 40mm, 50mm, 60mm, 90mm, 120mm, 150mm
Affected:All Lots
Software version:
Recalling Organisation:Stryker New Zealand Ltd
515 Mount Wellington Highway
Mount Wellington
AUCKLAND 1060
Contact Information:0800 787 9287
Manufacturer:Stryker GmbH
Issue:The manufacturer has confirmed that cracking of the Hoffmann® II Carbon Connecting Rods may occur following cleaning and sterilisation. Contrary to the original product labelling, the part is not suitable for reprocessing, and cleaning and sterilisation should not be performed, as this may result in cracks developing.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/7/2026