Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 37075 |
| Brand Name: | Flow Family Anaesthesia System |
| Model: | Part number: 6677300 - Flow-i C30 and Part number: 6887700 - Flow-c |
| Affected: | Multiple serial numbers |
| Software version: | |
| Recalling Organisation: | Getinge Australia Pty Ltd (NZ) Unit D, Ground Floor, Building A, Millenium Centre 600 Great South Rd Ellerslie Auckland 1051 |
| Contact Information: | 09 272 9039 |
| Manufacturer: | Maquet Critical Care AB |
| Issue: | The manufacturer identified a problem that under certain conditions might prevent the device from performing as intended. In some cases, Flow family anesthesia machines have not delivered anesthetic agent when the vaporizer(s) were switched on. These events have only been observed during the start-up phase of a case. This problem may stop the intended functioning of the device, resulting in failure to deliver anesthetic agent. This may result in the patient receiving insufficient anesthesia and may lead to intraoperative awareness. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 14/7/2026 |