Detail of Recall

Type of Product:Device
Medsafe Reference:37075
Brand Name:Flow Family Anaesthesia System
Model:Part number: 6677300 - Flow-i C30 and Part number: 6887700 - Flow-c
Affected:Multiple serial numbers
Software version:
Recalling Organisation:Getinge Australia Pty Ltd (NZ)
Unit D, Ground Floor, Building A, Millenium Centre
600 Great South Rd
Ellerslie
Auckland 1051
Contact Information:09 272 9039
Manufacturer:Maquet Critical Care AB
Issue:The manufacturer identified a problem that under certain conditions might prevent the device from performing as intended. In some cases, Flow family anesthesia machines have not delivered anesthetic agent when the vaporizer(s) were switched on. These events have only been observed during the start-up phase of a case.
This problem may stop the intended functioning of the device, resulting in failure to deliver anesthetic agent. This may result in the patient receiving insufficient anesthesia and may lead to intraoperative awareness.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:14/7/2026