Detail of Recall

Type of Product:Device
Medsafe Reference:36985
Brand Name:Smiths Medical ASD (MMSP) HOTLINE Fluid Warming Set
Model:Hotline Fluid Warming Set
Affected:Lot number - 4293569, 4337725
Software version:
Recalling Organisation:Smiths Medical Australasia Pty Ltd
c/o KPMG, KPMG Centre
18 Viaduct Harbour Avenue
Auckland
Contact Information:0800 444 200
Manufacturer:Smiths Medical ASD Inc
Issue:The manufacturer became aware of an increased likelihood of leakage in specific Lots of L-70 HOTLINE Fluid Warming Sets due to potentially inadequate solvent application at the tubing to return connector bond. Two (2) Lots of L-70 HOTLINE Fluid Warming Sets have an increased likelihood of leakage due to this problem.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:2/7/2026