Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:36948
Brand Name:RIDA®Plex Respiratory Pathogens C 16 well Step 1
Model:
Affected:Affected Lot S3520206
Software version:
Recalling Organisation:AusDiagnostics NZ Ltd
156B Bush Road
Albany
Auckland 0632
Contact Information:
Manufacturer:AusDiagnostics Pty Ltd
Issue:An increase in background signal in the RSV-A/B detection channel for affected Lot S3520206 was identified. This effect may generate atypical RSV-A/B amplification profiles resulting in incorrect RSV-A/B Check or false Positive calls.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:30/6/2026
Hide menus