| Type of Product: | Device |
| Medsafe Reference: | 36948 |
| Brand Name: | RIDA®Plex Respiratory Pathogens C 16 well Step 1 |
| Model: | |
| Affected: | Affected Lot S3520206 |
| Software version: | |
| Recalling Organisation: | AusDiagnostics NZ Ltd 156B Bush Road Albany Auckland 0632 |
| Contact Information: | |
| Manufacturer: | AusDiagnostics Pty Ltd |
| Issue: | An increase in background signal in the RSV-A/B detection channel for affected Lot S3520206 was identified. This effect may generate atypical RSV-A/B amplification profiles resulting in incorrect RSV-A/B Check or false Positive calls. |
| Recall action type: | Recall |
| Recall action: | Product to be destroyed |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 30/6/2026 |