Detail of Recall

Type of Product:Device
Medsafe Reference:36929
Brand Name:Portable ventilator, electric
Model:Astal 100 150 ventilators
Affected:
Software version:
Recalling Organisation:ResMed (NZ) Ltd
266 Runciman Road
Pukekohe
Auckland
Contact Information:0800 737 635
Manufacturer:ResMed Pty Ltd
Issue:The manufacturer had been made aware that an issue affecting a subset of Astral 100 and Astral 150 ventilators. An internal electrical component (supercapacitor) may leak electrolyte over time. In rare cases, this may damage specific circuitry on the Ventilator's printed circuit board assembly (PCBA), causing the ventilator to inadvertently enter a fail-safe state
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Healthcare Professional
Date Commenced:25/6/2026