Type of Product:Device
Medsafe Reference:36922
Brand Name:Horus TMC
Model:HORUS TMC cup
Affected:LOT numbers: 2010965, 2016014, 2017024, 2022633, 2021733
Software version:
Recalling Organisation:Evolutis New Zealand Limited
Unit 19, 150 Cavendish Road
Casebrook
Christchurch 8051
Contact Information:
Manufacturer:Evolutis
Issue:The device size is engraved on the back of the insert. A deviation has been observed in the depth of this engraving, which poses a potential risk of the thickness being less than the approved thickness.
The potential consequence is premature wear of the insert. All new batches manufactured since May 2026 comply with the requirements (no size engraving). Affected product that has not been implanted has been destroyed and replaced with new the new batch references so you will not receive any affected batches moving forward.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Healthcare Professional
Date Commenced:29/6/2026