Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:36916
Brand Name:CS - EOPA 3D Arterial Cannula
Model:78222 and 78322
Affected:0232612916, 0233245850, 0233245860, 0232862787
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Brand Name:CS - EOPA Elongated One-Piece Arterial Cannulae
Model:77422, 77522, 77622, 77722
Affected:0233009922, 0233116991, 0232798265, 0232902803, 0232601516, 0232601586, 0232601591, 0233009922, 0233528727, 0232990983,0232709428, 0232990983, 0232862794,0232862798, 0233286414, 0232745092, 0232903291, 0232601549, 0232602178, 0232744945
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Issue:The manufacturer received reports of pinhole leaks in the wire-wound body of 22 Fr EOPA 3D® Arterial Cannulae and EOPA™ Elongated One-Piece Arterial Cannulae used during cardiopulmonary bypass procedures. The leaks were identified either prior to use or during active bypass flow.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/6/2026