Type of Product:Device
Medsafe Reference:36901
Brand Name:BD Connecta stopcock
Model:BD Connecta stopcock with BD Q-Syte luer access split septum, BD Connecta stopcock without extension tube, BD Connecta Stopcock with OFF tap without extension tube, BD Connecta Stopcock with extension tube, and injection value
Affected:All Lot numbers
Software version:
Recalling Organisation:Becton Dickinson Ltd
Unit 4 / 110 Mays Road
Onehunga
Auckland 1061
Contact Information:0800 572 468
Manufacturer:BD Switzerland Sarl
Issue:The manufacturer identified an issue with Port C of the BD Connecta Stopcocks when used with polypropylene luer lock syringes. In some cases, the connection may over-thread and continue to spin, which can be perceived by users as unstable. Testing confirmed that the connection remains secure with no leakage under clinically relevant conditions, and the issue does not occur when used with polycarbonate syringes or IV tubing. The behaviour is due to lower torque required to override the threading when polypropylene syringes are used. While no functional failure was identified, this may cause user confusion during use. The manufacturer is issuing this communication to reinforce appropriate usage recommendations.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:18/6/2026