Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 36901 |
| Brand Name: | BD Connecta stopcock |
| Model: | BD Connecta stopcock with BD Q-Syte luer access split septum, BD Connecta stopcock without extension tube, BD Connecta Stopcock with OFF tap without extension tube, BD Connecta Stopcock with extension tube, and injection value |
| Affected: | All Lot numbers |
| Software version: | |
| Recalling Organisation: | Becton Dickinson Ltd Unit 4 / 110 Mays Road Onehunga Auckland 1061 |
| Contact Information: | 0800 572 468 |
| Manufacturer: | BD Switzerland Sarl |
| Issue: | The manufacturer identified an issue with Port C of the BD Connecta Stopcocks when used with polypropylene luer lock syringes. In some cases, the connection may over-thread and continue to spin, which can be perceived by users as unstable. Testing confirmed that the connection remains secure with no leakage under clinically relevant conditions, and the issue does not occur when used with polycarbonate syringes or IV tubing. The behaviour is due to lower torque required to override the threading when polypropylene syringes are used. While no functional failure was identified, this may cause user confusion during use. The manufacturer is issuing this communication to reinforce appropriate usage recommendations. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 18/6/2026 |