Detail of Recall

Type of Product:Device
Medsafe Reference:36884
Brand Name:STAAR Surgical AG - EVO Visian ICL V4c (VICMO) standard
Model:Visian EVO ICL V4c (VICMO) standard
Affected:
Software version:
Recalling Organisation:Device Technologies New Zealand
47 Arrenway Drive
Albany
AUCKLAND 0632
Contact Information:09 913 2000
Manufacturer:Staar Surgical AG
Brand Name:STAAR Surgical AG - EVO Visian ICL V4c (VTICMO) toric
Model:Visian EVO ICL V4c (VTICMO) toric
Affected:
Software version:
Recalling Organisation:Device Technologies New Zealand
47 Arrenway Drive
Albany
AUCKLAND 0632
Contact Information:09 913 2000
Manufacturer:Staar Surgical AG
Brand Name:STAAR Surgical AG - EVO Visian + (VICM5) standard
Model:Visian EVO+ (VICM5) standard
Affected:
Software version:
Recalling Organisation:KARL STORZ Endoscopy New Zealand Limited
31 Morningside Drive
Mt Albert
AUCKLAND 1025
Contact Information:
Manufacturer:Karl Storz SE & Co KG
Brand Name:STAAR Surgical AG - EVO+ Visian ICL (VTICM5) toric
Model:Visian EVO+ (VTICM5) toric
Affected:
Software version:
Recalling Organisation:Device Technologies New Zealand
47 Arrenway Drive
Albany
AUCKLAND 0632
Contact Information:09 913 2000
Manufacturer:Staar Surgical AG
Brand Name:STAAR Surgical - VICH lens
Model:Visian ICL (VICH)
Affected:
Software version:
Recalling Organisation:Device Technologies New Zealand
47 Arrenway Drive
Albany
AUCKLAND 0632
Contact Information:09 913 2000
Manufacturer:Staar Surgical AG
Brand Name:STAAR Surgical - VTICH lens
Model:Visian Toric ICL (VTICH)
Affected:
Software version:
Recalling Organisation:Device Technologies New Zealand
47 Arrenway Drive
Albany
AUCKLAND 0632
Contact Information:09 913 2000
Manufacturer:Staar Surgical AG
Issue:The manufacturer has become aware of problems that may affect information displayed on the Implant Orientation Diagram (IOD) that is generated in Stella 2.0. The IOD is used for the implantation of Toric EVO/EVO+ Implantable Collamer Lenses (ICLs).
The identified problems are as follows:
1. Expected refraction, spherical equivalent, and/or model/version values may be incorrect on the generated IOD.
2. If the sphere value for a Toric lens reservation is null, Stella 2.0 will default to indicate a rotation value of 90° on the IOD.
3. Users' local office network firewall settings may block Stella 2.0 web browser functionality, impacting the IOD display.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/6/2026