Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:36877
Brand Name: Radiography 7000 M
Model:Model number - 712010
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems DMC GmbH
Issue:The manufacturer has identified additional software problems that may impact clinical workflow.
(1) Unintended Backward Movement: The manufacturer has become aware that the Radiography 7000 M system may continue to move a short distance after releasing the drive handle while reversing. This issue only occurs if the drive handle is released before the system comes to a full stop.
(2) Detector Pairing Issue: When the Radiography 7000 M system clock is changed or restored (due to day light saving or after reinstalling software), there is a potential for the console user interface and tube control handle (TCH) user interface to incorrectly show a previously paired detector as connected.
(3) Dose value not displayed after an exposure (eDAP): Following a system reboot, the electronic Dose Area Product (eDAP) module intermittently fails to initialize, resulting in missing dose values for exposures recorded by the system.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:10/6/2026