Detail of Recall

Type of Product:Device
Medsafe Reference:36873
Brand Name:Guedel Airway Intersurgical
Model:Product code - 1100050 – One-piece Guedel airway, size 00, ISO 5.0, blue and Product code - 1000035 One-piece Guedel airway, size 000, ISO 3.5, light green
Affected:All lot numbers with an expiry date in the range 2026-06 to 2031-02, as indicated on the product label.
Software version:
Recalling Organisation:InterMed Medical Ltd
71 Apollo Drive
Albany
AUCKLAND 1311
Contact Information:0800 333 444
Manufacturer:Intersurgical Ltd
Issue:During manufacture, some Guedel Airway devices have been identified with occluded or partially occluded tips.
The hazard arising from this manufacturing fault depends on the level of occlusion. Potential problems include, but are not
limited to, respiratory distress, laryngospasm, stridor, stenosis, oedema, neurological impairment, hypoxia,
hypoxemia and cyanosis and their consequences.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/6/2026