Detail of Recall

Type of Product:Device
Medsafe Reference:36872
Brand Name:UHI-4_High Flow Insufflator Unit
Model:UHI-4 High Flow Insufflation Unit
Affected:All afected Lot numbers
Software version:
Recalling Organisation:Olympus (NZ) Ltd
15D Paul Matthews Road
Albany
Auckland
Contact Information:0508 659 6787
Manufacturer:Olympus Medical Systems Corporation
Issue:The manufacturer identified an increase in UHI-4 front-panel blackout complaints, where the front-panel display turns off unexpectedly, and air delivery (insufflation) ceases. An investigation found these blackouts (fault code E03) were primarily caused by the pressure sensors drifting positively from zero over time; when the pressure difference between the two sensors exceeds 7 mmHg, the UHI-4 generates the E03 fault, the front panel shuts off, and insufflation stops.
is working to implement a long-term solution for front panel blackout. In the interim, the UHI-4 pressure sensors must receive an offset adjustment to reset the sensor to zero because of sensor drift. This adjustment can be completed onsite and does not require returning your unit. If your UHI-4 displays a front panel blackout (fault code E03) during use, you must either reboot the unit or switch to a backup insufflator.
Recall action type:Correction
Recall action:
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/6/2026