Detail of Recall

Type of Product:Device
Medsafe Reference:36870
Brand Name:Angiographic X-Ray Allura Xper FD20
Model:Allura Xper FD20 - Model - 722012
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Brand Name:MultiDiagnost-Eleva systems
Model:MultiDiagnost Eleva with Flat Detector - Model - 708038
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:The manufacturer has observed occurrences in which the hand switch button, used for activating exposure from the control room, did not fully release after being pressed due to mechanical resistance within the button mechanism.
The manufacturer has observed occurrences in which the hand switch button, used for activating exposure from the control room, did not fully release after being pressed due to mechanical resistance within the button mechanism.
The hand switch has three mechanical positions:
1) Neutral – the button is not pressed.
2) Prepare – the button is partially pressed (first actuation point), during which the system enters a preparation phase for exposure imaging. During this phase, imaging functionality is not available.
3) Expose – the button is fully pressed (second actuation point), activates exposure imaging.
Recall action type:Correction
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/6/2026