Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 36845 |
| Brand Name: | Hoffman 3 Universal Pin Chucks Fixation device |
| Model: | Chuck for Pins 3 Universal for Dia 4,5,6mm |
| Affected: | Lot numbers - 1544AB, AN4261, AP3761, AR4000, AU0794 |
| Software version: | |
| Recalling Organisation: | Stryker New Zealand Ltd 515 Mount Wellington Highway Mount Wellington AUCKLAND 1060 |
| Contact Information: | 0800 787 9287 |
| Manufacturer: | Stryker GmbH |
| Issue: | The manufacturer confirmed that certain lots of Hoffman 3 Universal Pin Chucks contained inhomogeneous coating with out-of-specification coating thickness and colour. This nonconformance was discovered via product complaints. The potential hazard associated with this nonconformance is that the parts are not fully functional. The heterogeneity of the Ni PTFE coating may accelerate degradation and compromise device performance. In addition, visible discoloration may lead users to perceive the device as unclean, contaminated, or containing debris, despite the discoloration being cosmetic only. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 26/5/2026 |