Detail of Recall

Type of Product:Device
Medsafe Reference:36845
Brand Name:Hoffman 3 Universal Pin Chucks Fixation device
Model:Chuck for Pins 3 Universal for Dia 4,5,6mm
Affected:Lot numbers - 1544AB, AN4261, AP3761, AR4000, AU0794
Software version:
Recalling Organisation:Stryker New Zealand Ltd
515 Mount Wellington Highway
Mount Wellington
AUCKLAND 1060
Contact Information:0800 787 9287
Manufacturer:Stryker GmbH
Issue:The manufacturer confirmed that certain lots of Hoffman 3 Universal Pin Chucks contained inhomogeneous coating with out-of-specification coating thickness and colour. This nonconformance was discovered via product complaints.
The potential hazard associated with this nonconformance is that the parts are not fully functional. The heterogeneity of the Ni PTFE coating may accelerate degradation and compromise device performance. In addition, visible discoloration may lead users to perceive the device as unclean, contaminated, or containing debris, despite the discoloration being cosmetic only.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:26/5/2026