Detail of Recall

Type of Product:Device
Medsafe Reference:36841
Brand Name:Getinge 88-5 Washer disinfector
Model:Getinge 88-5 Washer disinfector (all units with electrical
Affected:Serial nubers - WAA060034, WAA068106, W50038320, W50046233
Software version:
Recalling Organisation:Getinge Australia Pty Ltd (NZ)
Unit D, Ground Floor, Building A, Millenium Centre
600 Great South Rd
Ellerslie
Auckland 1051
Contact Information:09 272 9039
Manufacturer:Getinge Disinfection AB
Issue:The initial investigation determined that the device was found to overheat during operation if the calibration of the circulation pressure was not performed or had been performed incorrectly and the alarm for low circulation pressure was not trigged.
The information sent out in previous customer letter Medsafe Ref#35754 was a precautionary action that was taken by checking the Circulation Pressure value in Stand By mode, to verify it was 0.0 kPa or otherwise re-calibrate.

Further investigation determined that the device could overheat during operation if the following conditions occurred:
· the water circulation pressure is too low in combination with:
· the circulation pressure calibration not being performed or performed incorrectly and/or
· The low circulation pressure alarm is not triggered due to the alarm limit being changed by a technician.

This can result in the melting of insulation, cables, and hoses, and in some cases, may lead to smoke emission or fire.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:26/5/2026