| Type of Product: | Device |
| Medsafe Reference: | 36835 |
| Brand Name: | Xray fluoro angio stationary digital |
| Model: | ARTIS icono floor, ARTIS icono biplane, ARTIS icono ceiling delivered with software version VE40A and VE40B |
| Affected: | Catalogue - 11327600, 11328100, 11327700 |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Healthcare GmbH |
| Issue: | The manufacturer has identified that in rare situations, after increasing the detector lift height to its maximum or minimum, the system may lose its movement capability. A reboot may not reestablish movement control. This failure can be resolved only by qualified service personnel. If this problem occurs, the system will report a failure of movement control and request a service call. It will not be possible to move the stand other than by emergency salvage procedures. Manual unmotorised table movements as well as fluoroscopy and acquisition will be still available. This may result in a situation where it is necessary to cancel the clinical procedure or to continue or begin the procedure on an alternative system. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 22/5/2026 |