Detail of Recall

Type of Product:Device
Medsafe Reference:36821
Brand Name:Philips Azurion Systems Angiographic X-Ray
Model:Philips Azurion 7 M20 Systems - system code 722079
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:The manufacturer has identified that the Float Tabletop control on the control module of Azurion systems may experience mechanical wear over time. The Float Tabletop control is located on the control module at the table-side. When the control is actuated, the table brakes disengage, allowing adjustment of the longitudinal or transverse position of the table. Over time, mechanical wear may lead to detachment, cracking, or breakage of the Float Tabletop control, which may lead to reduced availability or unavailability of longitudinal and transverse tabletop movement, as well as the formation of sharp edges.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/5/2026