Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 36798 |
| Brand Name: | InZone Detachment System |
| Model: | |
| Affected: | Catalogue number - 4510095 Multiple Lot numbers |
| Software version: | |
| Recalling Organisation: | Stryker New Zealand Ltd 515 Mount Wellington Highway Mount Wellington AUCKLAND 1060 |
| Contact Information: | 0800 787 9287 |
| Manufacturer: | Stryker Neurovascular |
| Issue: | The manufacturer has identified a problem affecting certain lots of InZone Detachment System devices. Stryker Neurovascular has observed that some devices may experience premature battery drain. When this occurs, the devices may: Not power on; Power on with faint audible and visual indicators; Be unable to detach a coil as intended. A corrective manufacturing step has been implemented. Replacement devices are available. |
| Recall action type: | Recall |
| Recall action: | Product to be exchanged |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 12/5/2026 |