Detail of Recall

Type of Product:Device
Medsafe Reference:36795
Brand Name:Merlin Patient Care System Model 3650
Model:MERLIN 3650 PATIENT CARE SYSTEM (PCS)
Affected:SOFTWARE MODEL 3330 V25.0.X – V28.7.X WHEN USED WITH AVEIR LEADLESS SYSTEM
Software version:
Recalling Organisation:Abbott Medical New Zealand Limited
4 Pacific Rise
Mt Wellingtion
Auckland 1640
Contact Information:0800 756 269
Manufacturer:St Jude Medical (USA)
Issue:The manufacturer is contacting customers to inform them of a Merlin PCS 3650 programmer software update that addresses a potential software behaviour observed during the Pacing Capture Threshold (PCT) test of the AVEIR™ Leadless Pacemaker (LP). In rare circumstances, if intermittent loss of telemetry occurs during a PCT test, the programmer may not successfully transmit the command to terminate the test, which could result in the LP remaining in sub-threshold pacing until telemetry is fully disconnected. This issue can only occur during a PCT test, which is performed at the time of implant or during in-clinic follow-up.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/5/2026