Type of Product:Device
Medsafe Reference:36787
Brand Name:MedChart
Model:
Affected:MedChart v12.6.1
Software version:
Recalling Organisation:DH (New Zealand) Limited
Level 2, 318 Lambton Quay
Wellington 6011
Contact Information:
Manufacturer:Dedalus (Australia) Pty Ltd
Issue:The manufacturer has identified that when calculating a dose from a QuickList/Protocol with the weight selected as Ideal Body Weight (IBW) or Adjusted Body Weight (AdjBW) for an adult (male or female), the dose is being calculated using the corresponding Child IBW or AdjBW weight formula instead of the adult calculator, resulting in a lower-than-expected dose.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory