Detail of Recall

Type of Product:Device
Medsafe Reference:36782
Brand Name:Nobel Biocare Implant Extractor
Model:Implant Retrieval Instrument CC RP & Tri-Channel WP 22mm
Affected:Lot number - 76719
Software version:
Recalling Organisation:Nobel Biocare New Zealand Ltd
Moore Markhams Auckland
Floor 1, 103 Carlton Gore Road
Newmarket
Auckland 1023
Contact Information:021 744 645
Manufacturer:Nobel Biocare AB
Issue:Due to a process error at the manufacturer, the Implant Retrieval Instrument CC RP & Tri-Channel WP 22mm has been mixed up with an Implant Retrieval instrument intended for the wide platform (WP). The laser marking on the device (CC RP) is in line with the packaging, but not correct for the delivered article.
The affected Implant Retrieval Instrument does not function as intended. Since the connecting part of the instrument is too wide, it cannot be used to retrieve a CC RP or Tri-Channel WP implant.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Healthcare Professional
Date Commenced:8/5/2026