Detail of Recall

Type of Product:Device
Medsafe Reference:36736
Brand Name:AnerysmFlow_Interventional Tool [MDR]
Model:Philips AneurysmFlow
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:The manufacturer has determined that the MAFA ratio does not provide reliable prognostic information regarding aneurysm occlusion following Flow Diverter Stent treatment. The Instructions for Use (IFU) of AneurysmFlow state that the MAFA ratio should not be used for making clinical decisions (Section 9.2 Measuring Flow). In addition, when hovering over the MAFA ratio header within the software interface, an on-screen message states that the MAFA ratio is not to be used for clinical decision making. Despite the existing cautions, the displayed MAFA ratio may still be considered during intra-procedural decision-making and could influence clinical judgment. This may potentially result in an incorrect clinical decision.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:16/4/2026