Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 36734 |
| Brand Name: | MEQU M Warmer |
| Model: | M Warmer |
| Affected: | Models - MWS201 (Box of 5), MWS201W (Single unit) |
| Software version: | |
| Recalling Organisation: | Oceania Medical Ltd 7/40 Arrenway Drive Rosedale Auckland 0632 |
| Contact Information: | |
| Manufacturer: | MEQU A/S |
| Issue: | The manufacturer has identified a manufacturing defect involving incorrect sealing, which may cause internal leakage and overheating of the device. This may result in: Activation of safety circuit (flashing red LED) Potential leakage of hot fluids Risk of patient skin burns if undetected |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 16/4/2026 |