Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:36734
Brand Name:MEQU M Warmer
Model:M Warmer
Affected:Models - MWS201 (Box of 5), MWS201W (Single unit)
Software version:
Recalling Organisation:Oceania Medical Ltd
7/40 Arrenway Drive
Rosedale
Auckland 0632
Contact Information:
Manufacturer:MEQU A/S
Issue:The manufacturer has identified a manufacturing defect involving incorrect sealing, which may cause internal leakage and overheating of the device.
This may result in:
Activation of safety circuit (flashing red LED)
Potential leakage of hot fluids
Risk of patient skin burns if undetected
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:16/4/2026
Hide menus
Show menus